dc.rights.license | http://creativecommons.org/licenses/by-nc/4.0 - Atribución-NoComercial | es_MX |
dc.contributor | J. GUADALUPE VELASQUEZ AGUILAR | es_MX |
dc.contributor.author | COSMOS CLEMENTE CATONGA | es_MX |
dc.contributor.other | director - Director | es_MX |
dc.coverage.spatial | MEX - México | es_MX |
dc.date | 2020-06-21 | |
dc.date.accessioned | 2020-06-21T01:20:55Z | |
dc.date.available | 2020-06-21T01:20:55Z | |
dc.identifier.uri | http://riaa.uaem.mx/handle/20.500.12055/1221 | |
dc.description | Resumen
En este trabajo, se desarrolló un generador de ondas Kotz (ondas rusas), que se puede utilizar
en la electroterapia de estimulación a frecuencia media. Para la realización del sistema, se hizo una
revisión de los conceptos elementales sobre las señales eléctricas y la forma en que estas funcionan
dentro del cuerpo humano. Los tipos de corrientes eléctricas y algunas de sus propiedades
específicas, son analizadas para identificar los efectos terapéuticos que pueden proporcionar al
aplicarse al cuerpo humano, en particular las ondas rusas.
Las señales generadas por el sistema son diseñadas utilizando modelos matemáticos, tanto
para la señal base (frecuencia constante) como para la señal moduladora (esquema de tratamiento en
base a tiempo, frecuencia y amplitud de la onda). Los datos obtenidos del modelado son
programados posteriormente en dispositivos electrónicos digitales reconfigurables, en este caso en
una matriz de puertas lógicas programables en campo (Field-Programmable Gate Array, FPGA).
Finalmente, el sistema en físico utiliza una tarjeta basada en FPGA modelo DE10-Lite. La
arquitectura de hardware se describe utilizando el VHDL (Lenguaje de Descripción de Hardware).
El sistema es validado mediante un banco de pruebas, generando diferentes señales y observando su
estabilidad en los períodos de tiempo de acuerdo con los utilizados en las terapias de rehabilitación. | es_MX |
dc.description | Abstract
In this work, a Kotz wave generator (Russian waves) was developed, which can be used in
medium frequency stimulation electrotherapy. To carry out the system, a review of the basic
concepts of electrical signals and the way they work within the human body was performed. The
types of electric currents and some of their specific properties are analyzed to identify the
therapeutic effects they can provide when applied to the human body, in particular Russian waves.
The signals generated by the system are designed using mathematical models, both for the
base signal (constant frequency) and for the modulation signal (treatment scheme based on time,
frequency and wave amplitude). The data obtained from the modeling is subsequently programmed
into reconfigurable digital electronic devices, in this case in a field programmable gate matrix
(FPGA) matrix.
Finally, the physical system uses a DE10-Lite FPGA-based card model. Hardware
architecture is described using VHDL (Hardware Description Language). The system is validated by
means of a test bench, which generates different signals and observes its stability over time in
accordance with those used in rehabilitation therapies. | es_MX |
dc.format | pdf - Adobe PDF | es_MX |
dc.language | spa - Español | es_MX |
dc.publisher | El autor | es_MX |
dc.rights | openAccess - Acceso Abierto | es_MX |
dc.subject | 7 - INGENIERÍA Y TECNOLOGÍA | es_MX |
dc.subject.other | 33 - CIENCIAS TECNOLÓGICAS | es_MX |
dc.title | Generador de ondas Kotz para electroestimulación muscular, basado en FPGA | es_MX |
dc.type | masterThesis - Tesis de maestría | es_MX |
uaem.unidad | Instituto de Investigación en Ciencias Básicas y Aplicadas (IICBA) - Instituto de Investigación en Ciencias Básicas y Aplicadas (IICBA) | es_MX |
uaem.programa | Maestría en Ingeniería y Ciencias Aplicadas - Maestría en Ingeniería y Ciencias Aplicadas | es_MX |
dc.type.publication | acceptedVersion | es_MX |
dc.audience | researchers - Investigadores | es_MX |